The RetroSys™ RT Activity Kit
Product code: RT-001
Protocol for screening for RT activity in cell supernatants.
80 samples set at two assay durations may be analysed with one kit.
- Collect kit components.
- Take out the following components from the refrigerator:
2 PolyA Plates
1 Sample Dilution Buffer (bottle B)
2 RT Reaction Components (vial C1)
1 Reconstitution Buffer (bottle C2)
1 HIV-1 rRT-standard (vial D)
1 Concentrated Washing Buffer (bottle E)
4 pieces of adhesive tape
- Prepare the reaction mixture.
- Add 12 ml of Reconstitution Buffer (C2) to each of the two vials
containing RT Reaction Components (C1). Make sure that the lyophilised
material dissolves properly.
- Transfer the contents of both C1 vials to a 50 ml glass beaker.
- Add 12 ml of distilled water.
- Add 12 ml of Sample Dilution Buffer and mix thoroughly
- Prepare the PolyA Plates.
- Take out both PolyA Plates from their foil pouches. Save one foil
pouch for later use.
- Add 200 µl of reaction mixture (beaker) to each well of the plates.
- Seal the plates with adhesive tape.
- Incubate the plates at 33°C for 20-60 minutes
- Dilute the HIV-1 rRT-standard.
When preparing the standard dilutions, mix thoroughly and change
pipette tips between each transfer.
- Add 1.2 ml of Sample Dilution Buffer (B) to the vial containing
lyophilised HIV-1 rRT-standard (D). Make sure that the lyophilised
material dissolves properly and mix thoroughly.
- Add 250 µl of Sample Dilution Buffer (B) to twelve test tubes.
- Add 100 µl of HIV-1 rRT standard (D) to the first of the twelve
test tubes.
- Transfer 200 µl from the first test tube to the second.
- Transfer 200 µl from the second test tube to the third and so on
until a series of twelve dilutions has been obtained.
- Prepare the samples.
Avoid aspirating cells when sampling cell culture supernatants.
- Collect the samples to be analysed. 2 x 10 µl of each sample is
required.
- Start the RT reaction.
When adding samples to the plates, change pipette tips between each
sample.
- Take out the PolyA Plates from the incubator.
- Add 10 µl of the first sample to well A1 of both PolyA Plates.
- Add 10 µl of the second sample to well B1 and so on until all 80
samples are set on both plates.
- Add 10 µl of Sample Dilution Buffer (B) to wells E12-H12 of both
plates.
- Add 10 µl of the first standard dilution to well A11 of both plates.
- Add 10 µl of the second standard dilution to well B11 and so on
until all twelve standard dilutions are set on both plates.
- Seal the plates with adhesive tape. Press the tape down firmly over
each well to ensure proper sealing.
- Incubate the plates at 33°C on an orbital shaker set at gentle agitation.
The first plate is incubated for 3 hours and the second overnight.
- Prepare the wash fluid.
- Add 75 ml of Triton X-100 to 1 litre of distilled water and
mix thoroughly with a magnetic spin bar.
Make sure that the Triton X-100 dissolves completely which takes
approximately 10 minutes.
- Add 10 ml of Concentrated Washing Buffer (E) to a 10 litre container.
- Add the dissolved Triton X-100 to the 10 litre container.
- Adjust the volume to 10 litres with distilled water and mix thoroughly.
The residual Concentrated Washing Buffer (E) is used on day two
and should be stored in the refrigerator until then.
- Stop the RT reaction of the first plate.
After 3 hours of incubation the RT reaction of the first plate is
stopped by washing. Use approximately 6 ml of wash fluid per well and
wash.
- Wash the PolyA Plate twice. Turn the plate between washes if possible.
- Remove residual fluid in the wells by tapping the plate upside down
on absorbing cloth or paper.
- Return the plate to its foil pouch and store it at -20°C.
Day two
- Collect kit components.
- Take out the kit box from the refrigerator and take out the following
components:
1 Concentrated Washing Buffer (bottle E)
2 RT Product Tracer (vials O)
1 AP Substrate Tablets (tube P1)
1 AP Substrate Buffer (bottle P2)
2 pieces of adhesive tape
2 pieces of plastic lids
- Prepare the wash fluid.
Follow the procedure for this as described in step 7.
- Prepare the RT Product Tracer.
- Add 12 ml of distilled water to each of the two vials containing
RT Product Tracer (O). Make sure that the lyophilised material dissolves
properly and mix thoroughly.
- Transfer the contents of both O vials to a 50 ml glass beaker.
- Stop the RT reaction of the second plate.
After overnight incubation the RT reaction of the second plate is stopped
by washing. Use approximately 6 ml per well and wash.
- Wash the PolyA Plate twice. Turn the plate 180° between washes
if possible.
- Remove residual fluid in the wells by tapping the plate upside
down on absorbing cloth or paper.
- Collect the first PolyA Plate from the freezer.
- Start the binding of RT Product Tracer.
When dispensing the tracer, pipette tips should not touch the bottom
of the wells.
- Add 100 µl to each well of both PolyA Plates.
- Seal the plates with adhesive tape. Press the tape down firmly
over each well to ensure proper sealing.
- Incubate the plates for 90 minutes at 33°C on an orbital shaker
set at gentle agitation.
- Prepare the AP substrate solution.
- Add the AP Substrate Tablets (P1) to the AP Substrate Buffer
(P2).
- Shake the bottle occasionally and the tablets will dissolve completely
in approximately twenty minutes.
- Remove excess tracer.
After 90 minutes of incubation, excess tracer is removed by washing.
Use approximately 6 ml of wash fluid per well and wash.
- Wash both PolyA Plates four times. Turn the plates 180° between
washes if possible.
- Remove residual fluid in the wells by tapping the plate upside
down on absorbing cloth or paper.
- Start the alkaline phosphatase reaction.
The AP substrate solution must be mixed thoroughly and equilibrated
to room temperature before it is added to the PolyA Plates. When dispensing
the substrate solution, pipette tips should not touch the bottom of
the wells.
- Add 200 µl of AP substrate solution to each well of both PolyA
Plates.
- Cover the plates with plastic lids and incubate them at room
temperature under dark cover on an orbital shaker set at gentle agitation.
- Read the plates.
Set the filter of the plate reader to 405 nm.
- Read the absorbance of both plates after 30 minutes of incubation.
- Read the absorbance of both plates after 2 hours of incubation.
- Read the absorbance of both plates after overnight incubation.
- Process the data.
The RetroSys RT assay gives a linear relationship between A405
and RT activity if the readings are within the linear measuring
range of the plate reader used
When calculating the results the data from each reading time should
be processed separately, i.e. values for standard curve and samples
should be from the same reading time.
- Plot the A405 of each standard dilution (wells A11-D12)
against its concentration of HIV-1 rRT (see table below). Preferentially
use a computer program to calculate the best regression line utilising
the mean A405 of the background controls (wells E12-H12)
as y-axis intercept. Only use readings within the linear measuring
range of the microtiter plate reader.
- Determine the RT activity of each well with the aid of the regression
line. Only determine RT activity for wells giving an A405 within
the linear measuring range of the reader and greater than two times
the mean of the background controls.
- Compare RT activity values for long and short RT assay incubations
for each sample.
a) If the RT values are similar then calculate the mean value
for the sample.
b) If a higher value is obtained for the short RT reaction time
then this is the value of the sample. The lower value obtained with
the long RT reaction time is normally due to RT product degradation.
c) If a significant value only is obtained for the long RT reaction
time then this is the value of the sample.
For samples where the first colour reading gives an A405
above the linear measuring range of the plate reader, the RT activity
is higher than the highest HIV-1 rRT activity included in the standard
curve. If an accurate value for such a sample is desired, the sample
should be re-analysed using several dilutions of the sample (see Protocol
for quantification of RT activity).
- Take out the following components from the refrigerator:
2 PolyA Plates
1 Sample Dilution Buffer (bottle B)
2 RT Reaction Components (vial C1)
1 Reconstitution Buffer (bottle C2)
1 HIV-1 rRT-standard (vial D)
1 Concentrated Washing Buffer (bottle E)
4 pieces of adhesive tape
- Add 12 ml of Reconstitution Buffer (C2) to each of the two vials containing RT Reaction Components (C1). Make sure that the lyophilised material dissolves properly.
- Transfer the contents of both C1 vials to a 50 ml glass beaker.
- Add 12 ml of distilled water.
- Add 12 ml of Sample Dilution Buffer and mix thoroughly
- Take out both PolyA Plates from their foil pouches. Save one foil pouch for later use.
- Add 200 µl of reaction mixture (beaker) to each well of the plates.
- Seal the plates with adhesive tape.
- Incubate the plates at 33°C for 20-60 minutes
When preparing the standard dilutions, mix thoroughly and change pipette tips between each transfer.
- Add 1.2 ml of Sample Dilution Buffer (B) to the vial containing lyophilised HIV-1 rRT-standard (D). Make sure that the lyophilised material dissolves properly and mix thoroughly.
- Add 250 µl of Sample Dilution Buffer (B) to twelve test tubes.
- Add 100 µl of HIV-1 rRT standard (D) to the first of the twelve test tubes.
- Transfer 200 µl from the first test tube to the second.
- Transfer 200 µl from the second test tube to the third and so on
until a series of twelve dilutions has been obtained.
Avoid aspirating cells when sampling cell culture supernatants.
- Collect the samples to be analysed. 2 x 10 µl of each sample is
required.
When adding samples to the plates, change pipette tips between each sample.
- Take out the PolyA Plates from the incubator.
- Add 10 µl of the first sample to well A1 of both PolyA Plates.
- Add 10 µl of the second sample to well B1 and so on until all 80 samples are set on both plates.
- Add 10 µl of Sample Dilution Buffer (B) to wells E12-H12 of both plates.
- Add 10 µl of the first standard dilution to well A11 of both plates.
- Add 10 µl of the second standard dilution to well B11 and so on until all twelve standard dilutions are set on both plates.
- Seal the plates with adhesive tape. Press the tape down firmly over each well to ensure proper sealing.
- Incubate the plates at 33°C on an orbital shaker set at gentle agitation.
The first plate is incubated for 3 hours and the second overnight.
- Add 75 ml of Triton X-100 to 1 litre of distilled water and
mix thoroughly with a magnetic spin bar.
Make sure that the Triton X-100 dissolves completely which takes approximately 10 minutes. - Add 10 ml of Concentrated Washing Buffer (E) to a 10 litre container.
- Add the dissolved Triton X-100 to the 10 litre container.
- Adjust the volume to 10 litres with distilled water and mix thoroughly.
The residual Concentrated Washing Buffer (E) is used on day two and should be stored in the refrigerator until then.
After 3 hours of incubation the RT reaction of the first plate is stopped by washing. Use approximately 6 ml of wash fluid per well and wash.
- Wash the PolyA Plate twice. Turn the plate between washes if possible.
- Remove residual fluid in the wells by tapping the plate upside down on absorbing cloth or paper.
- Return the plate to its foil pouch and store it at -20°C.
Day two
- Take out the kit box from the refrigerator and take out the following
components:
1 Concentrated Washing Buffer (bottle E)
2 RT Product Tracer (vials O)
1 AP Substrate Tablets (tube P1)
1 AP Substrate Buffer (bottle P2)
2 pieces of adhesive tape
2 pieces of plastic lids
Follow the procedure for this as described in step 7.
- Add 12 ml of distilled water to each of the two vials containing RT Product Tracer (O). Make sure that the lyophilised material dissolves properly and mix thoroughly.
- Transfer the contents of both O vials to a 50 ml glass beaker.
After overnight incubation the RT reaction of the second plate is stopped by washing. Use approximately 6 ml per well and wash.
- Wash the PolyA Plate twice. Turn the plate 180° between washes if possible.
- Remove residual fluid in the wells by tapping the plate upside down on absorbing cloth or paper.
- Collect the first PolyA Plate from the freezer.
When dispensing the tracer, pipette tips should not touch the bottom of the wells.
- Add 100 µl to each well of both PolyA Plates.
- Seal the plates with adhesive tape. Press the tape down firmly over each well to ensure proper sealing.
- Incubate the plates for 90 minutes at 33°C on an orbital shaker
set at gentle agitation.
- Add the AP Substrate Tablets (P1) to the AP Substrate Buffer (P2).
- Shake the bottle occasionally and the tablets will dissolve completely
in approximately twenty minutes.
After 90 minutes of incubation, excess tracer is removed by washing. Use approximately 6 ml of wash fluid per well and wash.
- Wash both PolyA Plates four times. Turn the plates 180° between washes if possible.
- Remove residual fluid in the wells by tapping the plate upside
down on absorbing cloth or paper.
The AP substrate solution must be mixed thoroughly and equilibrated to room temperature before it is added to the PolyA Plates. When dispensing the substrate solution, pipette tips should not touch the bottom of the wells.
- Add 200 µl of AP substrate solution to each well of both PolyA Plates.
- Cover the plates with plastic lids and incubate them at room
temperature under dark cover on an orbital shaker set at gentle agitation.
Set the filter of the plate reader to 405 nm.
- Read the absorbance of both plates after 30 minutes of incubation.
- Read the absorbance of both plates after 2 hours of incubation.
- Read the absorbance of both plates after overnight incubation.
The RetroSys RT assay gives a linear relationship between A405 and RT activity if the readings are within the linear measuring range of the plate reader used
When calculating the results the data from each reading time should be processed separately, i.e. values for standard curve and samples should be from the same reading time.
- Plot the A405 of each standard dilution (wells A11-D12) against its concentration of HIV-1 rRT (see table below). Preferentially use a computer program to calculate the best regression line utilising the mean A405 of the background controls (wells E12-H12) as y-axis intercept. Only use readings within the linear measuring range of the microtiter plate reader.
- Determine the RT activity of each well with the aid of the regression line. Only determine RT activity for wells giving an A405 within the linear measuring range of the reader and greater than two times the mean of the background controls.
- Compare RT activity values for long and short RT assay incubations
for each sample.
a) If the RT values are similar then calculate the mean value for the sample.
b) If a higher value is obtained for the short RT reaction time then this is the value of the sample. The lower value obtained with the long RT reaction time is normally due to RT product degradation.
c) If a significant value only is obtained for the long RT reaction
time then this is the value of the sample.
For samples where the first colour reading gives an A405 above the linear measuring range of the plate reader, the RT activity is higher than the highest HIV-1 rRT activity included in the standard curve. If an accurate value for such a sample is desired, the sample should be re-analysed using several dilutions of the sample (see Protocol for quantification of RT activity).